172 • Configuring the PID Gas in Setup Mode

EAGLE 2 Operator’s Manual

12. Insert the sensor adapter into the PID flow chamber and push it in until it

bottoms out.

13. Plug the PID cable connector into the PID sub PCB on the main PCB.

14. Confirm that the main PCB is seated in its slots and that its bottom edge

is resting on the bottom of the bottom case. If the main PCB is not seated
properly, it may be damaged when the top case is re-installed.

15. Make sure that the top case gasket is fully seated in its groove and

carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.

16. Turn the EAGLE 2 upside down and tighten the three case screws to

secure the top case to the bottom case.

17. Calibrate the PID channel as described in “PID Calibration” on page 143.

Configuring the PID Gas in
Setup Mode

The standard PID channel is configured for and calibrated to isobutylene. If
calibration to a different gas is required for an application, the PID channel
can also be configured for other gases in the CONFIGURE GASES menu
item in Setup Mode. To change the gas configuration of the PID channel in
Setup Mode, do the following:

WARNING: The EAGLE 2 is not in operation as a gas detector while in

Setup Mode.

1. Take the EAGLE 2 to a non-hazardous location and turn it off if it is on.

2. Press and hold AIR ▲ YES and RANGE ▼ SHIFT, then press and hold

POWER ENTER RESET. When you hear a beep, release the buttons.

3. The LCD will show the following screen for a few seconds with the “S”

in the lower right corner indicating the unit is entering Setup Mode.

GAS MONITOR

EAGLE 2 S