26 • Start Up

EAGLE 2 Operator’s Manual

will again appear in the startup sequence.

•

If the unit is due for calibration and CAL PAST DUE ACT is set to
NOTIFICATION ONLY, then the following alert screen displays and
the buzzer sounds in a double pulsing pattern.

Press and release the POWER ENTER RESET button to
acknowledge the alert and continue with the startup sequence.

10. The Date/Time Screen appears for a few seconds.

11. If USER/STATION ID is turned on (see “Turning the User/Station ID

Function On or Off” on page 118), the ID Screen appears for a few
seconds.

If USER/STATION ID is turned off, only the serial number is shown.

12. If the EAGLE 2 experiences a sensor failure during start up, a screen

indicating which sensor failed appears and the buzzer sounds a pulsing
tone twice per second. In the example below, the H

2

S sensor has failed.

CALIBRATION DATE

IS PAST DUE

9/12/2008

15:00:00

USER ID

MIKE

STATION ID

PUMP 1

SERIAL NUMBER

E2A515

FAILED SENSOR(S)

< > < >

<H2S> < >

ENTER TO CONTINUE