26 • Start Up

EAGLE 2 Operator’s Manual

will again appear in the startup sequence.

If the unit is due for calibration and CAL PAST DUE ACT is set to
NOTIFICATION ONLY, then the following alert screen displays and
the buzzer sounds in a double pulsing pattern.

Press and release the POWER ENTER RESET button to
acknowledge the alert and continue with the startup sequence.

10. The Date/Time Screen appears for a few seconds.

11. If USER/STATION ID is turned on (see “Turning the User/Station ID

Function On or Off” on page 118), the ID Screen appears for a few
seconds.

If USER/STATION ID is turned off, only the serial number is shown.

12. If the EAGLE 2 experiences a sensor failure during start up, a screen

indicating which sensor failed appears and the buzzer sounds a pulsing
tone twice per second. In the example below, the H

2

S sensor has failed.

CALIBRATION DATE

IS PAST DUE

9/12/2008

15:00:00

USER ID

MIKE

STATION ID

PUMP 1

SERIAL NUMBER

E2A515

FAILED SENSOR(S)

< > < >

<H2S> < >

ENTER TO CONTINUE