24 • Start Up

EAGLE 2 Operator’s Manual

from the last time the EAGLE 2 was used. See “Time in Operation
Screen” on page 51 for more information about how the EAGLE 2
tracks the operating time. The short-term exposure limit (STEL)
reading is reset each time the EAGLE 2 is turned on.

To reset the accumulation of these measurements, press and release
the DISPLAY ADJUST NO button before the countdown reaches 0.

8. The gas alarm setpoints are displayed by three screens in sequence: the

Low Alarm Screen, High Alarm Screen, and STEL/TWA Alarm Screen.
Each screen remains on the LCD for three seconds.

9. After the alarm screens, if CAL REMINDER is turned on, the screen that

appears next depends on how CAL PAST DUE ACT is set in the Setup
Mode Menu (see “Updating the Calibration Past Due Action Setting” on
page 123).

A CH4 10 %LEL

L L OXY 19.5 vol%

O A H2S 10.0 ppm

WR CO 25 ppm

M

S

CH4 10 vol%

A CH4 50 %LEL

HL OXY 23.5 vol%

I A H2S 30.0 ppm

GR CO 50 ppm

HM

S

CH4 50 vol%

ALARMS STEL & TWA

H2S(ppm) 15.0 10.0

CO (ppm) 200 25