EAGLE 2 Operator’s Manual

Using Setup Mode • 119

2. Press and release POWER ENTER RESET. The Adjust Auto Calibration

Values Screen appears. The auto calibration value for each channel is
shown.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to place the cursor next to the auto

calibration value you want to change.

4. Press and release POWER ENTER RESET. The auto calibration value

begins to flash indicating it can be adjusted.

5. Use AIR ▲ YES and RANGE ▼ SHIFT to adjust the auto calibration

value to the desired value.

6. Press and release the POWER ENTER RESET button to accept the

value.

7. Repeat Step 3 - Step 6 for each auto calibration value you want to change.

If you want to return to the main menu at any time without saving any
changes, press and release the DISPLAY ADJUST NO button until you
return to the main menu.

8. Use RANGE ▼ SHIFT to move the cursor next to the END menu item.

Press and release the POWER ENTER RESET button to save the
changes and return to the main menu.

Updating the Backlight Delay Setting

This setting indicates the length of time the LCD illuminates when you press
any button. The minimum setting is 0 seconds; the maximum setting is 255
seconds. The factory setting is 30 seconds.

1. From the main menu, place the cursor next to BACKLIGHT DELAY.

2. Press and release POWER ENTER RESET. The Backlight Delay Screen

appears.

ADJUST AUTO

CALIBRATION VALUES

> CH4 50 %LEL

OXY 12.0 vol%
H2S 25.0 ppm

CO 50 ppm

BACKLIGHT DELAY

UP/DOWN THEN ENTER

30 SECONDS