118 • Using Setup Mode

EAGLE 2 Operator’s Manual

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Turning the User/Station ID Function On or Off

With USER/STATION ID set to ON, the ID Screen displays during start up
and the Select User ID Screen and Select Station ID Screen appear in Display
Mode. The ID’s can be selected in Display Mode.

With USER/STATION ID set to OFF (factory setting), the ID Screen does
not display during start up and the Select User ID Screen and Select Station
ID Screen do not appear in Display Mode.

1. From the main menu, place the cursor next to USER/STATION ID.

2. Press and release POWER ENTER RESET. The User and Station ID’s

Screen appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating the Autocal Values

The EAGLE 2 stores calibration gas value settings. This allows you to
calibrate all EAGLE 2 channels simultaneously with a calibration cylinder
that contains all required target gases (for example the RKI 4-gas calibration
cylinder).

The EAGLE 2 includes default auto calibration settings for most target gases.
For gases without a default auto calibration value, the setting is 0.

NOTE: You can also update auto calibration settings in Calibration Mode.

Updating the auto calibration gas values in Calibration Mode is
normally done when performing a calibration. Updating these
settings in Setup Mode allows you to update the settings without
performing a calibration.

1. From the main menu, place the cursor next to ADJ AUTOCAL

VALUES.

USER AND

STATION ID’S

UP/DOWN THEN ENTER

OFF