RKI Instruments EAGLE 2 Manual User Manual

Page 40

Advertising
background image

30 • Measuring Mode, Normal Operation

EAGLE 2 Operator’s Manual

Monitoring an Area

1. Start up the EAGLE 2 as described above in “Start Up” on page 22. It is

now in Measuring Mode.

2. Take the EAGLE 2 to the monitoring area.

Put the probe tip in the area to be monitored.

NOTE: If you notice that water has entered the flow system through the

probe, replace the probe’s hydrophobic filter and inspect the O-ring
for filter particles. See “Replacing the Hydrophobic Probe’s Particle
Filter and Hydrophobic Filter Disk” on page 75 for instructions to
replace the hydrophobic filter.

3. Wait 10 - 15 seconds and observe the display for gas readings. If a

reading is observed, allow the reading to stabilize to determine the gas
concentrations present.

NOTE: Response time increases with the length of the sample hose. Long

sample hoses will require more time to show a response at the
EAGLE 2. The maximum sample hose length recommended for the
EAGLE 2 is 125 feet. Consult RKI Instruments, Inc. for longer
sample hose lengths.

4. If a gas alarm occurs, take appropriate action. See “Responding to

Alarms” on page 38.

Using Optional Sample Hoses

The standard sample hose for the EAGLE 2 is 5 feet long. Optional hoses are
available up to 125 feet long. If you are considering using a longer hose, keep
in mind that a longer hose will increase the EAGLE 2’s response time and the
flowrate may decrease close to the low flow alarm point.

CAUTION: Sample hose lengths of more than 125 feet are not recommended

for the EAGLE 2 because of flow rate reduction and increased
response time. Consult RKI Instruments, Inc. for hose lengths
longer than 125 feet.

CH4 0%LEL

OXY 20.9vol%

H2S 0.0ppm

CO 0ppm

Advertising