Start Up • 297

WARNING: The EAGLE 2 is not a gas monitoring device until you select

a mode and enter it.

5. If you select Normal Mode, the warm up sequence will continue as

described in “Start Up” on page 22 unless Leak Check or Bar Hole Mode
is active. If Leak Check or Bar Hole Mode is active, the Leak Check/Bar
Hole Mode Select Screen will appear immediately after the Normal
Mode selection.

Use the AIR ▲ YES or RANGE

SHIFT button to display the mode you

wish to enter.

6. If you select Inert Mode, the warm up sequence will continue as

described in “Start Up” on page 22. The oxygen alarms in the alarm point
screens will be the alarm points for Inert Mode and not for Normal Mode.

7. When the warm up is complete and the instrument is operating in Inert

Mode, the screen will appear as shown below if the instrument is
sampling a fresh air environment.

If the instrument is sampling a fresh air environment, the oxygen channel
will be in alarm.

> NORMAL MODE

BAR HOLE MODE

LEAK CHECK MODE

I CH4 0%LEL

N OXY 20.9vol%ALRM2

E H2S 0.0ppm

R CO 0ppm