RKI Instruments EAGLE 2 Manual User Manual

Page 175

Advertising
background image

EAGLE 2 Operator’s Manual

Maintenance • 165

28. Confirm that the main PCB is seated in its slots and that its bottom edge

is resting on the bottom of the bottom case. If the main PCB is not seated
properly, it may be damaged when the top case is re-installed.

29. Make sure that the top case gasket is fully seated in its groove and

carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.

30. Turn the EAGLE 2 upside down and tighten the three case screws to

secure the top case to the bottom case.

31. Calibrate the PID channel as described in “PID Calibration” on page 143.

Replacing the PID Sensor’s Lamp

If cleaning the PID lamp does not resolve any calibration problems you may
be having, the lamp may need to be replaced. Perform the following
procedure to replace the PID lamp:

NOTE: Do not touch the new lamp window (the flat end) with your fingers

as this may contaminate the window with finger oil.

1. Verify that the EAGLE 2 is off.

2. Place the EAGLE 2 upside down on a flat surface or hold it upside down.

3. Unscrew the three case screws until they disengage from the top case.

They are captive screws so they will not fall off of the bottom case.

4. Turn the EAGLE 2 right side up and carefully lift the top case away from

the bottom case. Be careful not to lift it so far that it pulls on the main
PCB with the cable that connects the top case to the main PCB.

5. Lay the top case down next to the bottom case to allow access to the flow

system.

6. Locate the PID sensor adapter. It has a five wire cable and is normally

located next to the pump. The cable has a connector that mates to a PID
sub PCB that is installed on the main PCB. Figure 25 on page 137 shows
a PID sensor in a typical location.

7. Grasp the sensor adapter firmly and pull it out of the PID flow chamber.

Rock it back and forth gently if necessary to pull it out. Take care not to
pull the cable connector from the PID sub PCB.

8. The PID sensor protrudes from one end of the sensor adapter. Grasp the

PID sensor firmly and pull it out of the sensor adapter.

9. Place the PID sensor face down on a flat clean working surface.

10. Hold the PID sensor steady on the working surface with one hand and

Advertising