EAGLE 2 Operator’s Manual

ESM-01 Calibration • 197

29. If you want to change the ESM-01 channel’s calibration gas value, follow

Step 15 - Step 24 above beginning with the ESM-01 Calibration Gas
Value Screen below instead of the standard channel Calibration Gas
Value Screen.

30. With the ESM-01 Calibration Gas Value Screen displayed, press and

release the POWER ENTER RESET button to proceed to the Calibration
In Process Screen for the ESM-01 channel with CAL IN PROCESS
flashing.

If you do not want to proceed with the calibration, press and release the
DISPLAY ADJUST NO button to return to the ESM-01 Cal Gas Values
Screen.

If you do want to continue with the calibration, proceed to the next step.

31. Since calibration gas has already been flowing to the ESM-01 sensor

while the other channels were being calibrated, the ESM-01 sensor
reading should be stable and ready for adjustment.

32. Press and release the POWER ENTER RESET button to set the span

adjustment for the ESM-01 channel to the programmed value.

CAL GAS VALUE

SO2 5 ppm

ENTER TO BEGIN CAL

CAL IN PROCESS

SO2 3 ppm

ENTER WHEN DONE