122 • Using Setup Mode

EAGLE 2 Operator’s Manual

shows the remaining log time and does not give you the opportunity to clear
the logged data.

1. From the main menu, place the cursor next to DATA LOG MEMORY.

2. Press and release POWER ENTER RESET. The Prompt to Clear Data

Log Memory? Screen appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating the LCD Contrast Setting

The contrast setting controls the LCD contrast. Select the setting so the
characters on the display are easy to see. It can be set from 1 to 15. The
factory setting is 8. The higher the setting, the darker the characters and LCD
background.

1. From the main menu, place the cursor next to ADJUST CONTRAST.

2. Press and release POWER ENTER RESET. The Adjust Contrast Screen

appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to adjust the setting so that the

characters on the LCD are easy to see.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating the Calibration Reminder Setting

With CAL REMINDER set to ON (factory setting), the EAGLE 2 will give
an indication at start up if it is due for calibration. The type of indication will
depend on the CAL PAST DUE ACT setting (see the next menu item
below).

PROMPT TO CLEAR
DATA LOG MEMORY?

UP/DOWN THEN ENTER

ON

ADJUST CONTRAST

UP/DOWN THEN ENTER

8