268 • Monitoring in Methane Elimination Mode

EAGLE 2 Operator’s Manual

WARNING: The EAGLE 2’s alarms are initiated by the display reading,

not the factored reading. If you are monitoring for isobutane
as in the above example and the low alarm is set for 10%
LEL, the EAGLE 2 will initiate a low alarm at 5.9% LEL
isobutane (display reading of 10% LEL).

12. To return to full response mode, return to the Methane Elimination Mode

screen in Display Mode and press and release the AIR ▲ YES or
RANGE

SHIFT button to toggle the setting to DISABLED.

13. Press and release the POWER ENTER RESET button. The unit will save

the setting and proceed to the next menu item.

14. Continue to press the DISPLAY button until the Measuring Mode screen

appears.

The ME is no longer next to the catalytic combustible channel reading

1-Butyl Alcohol

1.71

2.18

Methyl Acrylate

0.83

2.11

Cyclohexane

1.06

1.25

Methyl Ethyl
Keytone

1.04

1.32

Cumene

1.74

1.42

Methyl Isobutyl
Keytone

1.30

1.42

Ethylene
Dichloride

2.04

11.50

Mixed Xylenes

1.36

1.36

Ethyl Alcohol

0.57

1.71

Nonane

1.66

1.21

Ethyl Chloride

0.59

2.04

Pentane

0.54

0.74

Ethyl Acrylate

1.32

1.68

Propane

Low

response

Low

response

Hexane

1.00

1.00

Styrene

1.74

1.42

Hydrogen

0.48

1.75

Toluene

1.25

1.25

Vinyl Acetate
Monomer

0.92

2.17

*

Vapor pressure too low for significant LEL reading

Target Gas

LEL Factor

PPM Factor

Target Gas

LEL Factor

PPM Factor

HEX 0%LEL

OXY 20.9vol%

H2S 0.0ppm

CO 0ppm

Advertising