220 • Configuring the TC Gas in Setup Mode

EAGLE 2 Operator’s Manual

WARNING: Do not use the user defined gas configuration without

consulting RKI Instruments, Inc.

14. Enter the gas name. There are 3 characters available for the gas name.

The factory setting of TC uses only 2 characters. Use AIR ▲ YES and
RANGE ▼ SHIFT to display the desired character, then press POWER
ENTER RESET to enter the displayed character and move to the next
character. Repeat until all three characters are entered. When the last
character is entered, the full scale value will be flashing.

15. Use AIR ▲ YES and RANGE ▼ SHIFT to display the desired full scale

value. It may be any value from 10 to 100 in increments of 10.

16. Press and release POWER ENTER RESET to enter the full scale value.

The confirmation screen appears. In the example below, the user defined
gas has been selected and defined as propane with the gas name set to
PRO.

17. If you want to accept the gas configuration change, press and release AIR

▲ YES. The unit will return to the Configure Gases screen.

If you do not want to accept the gas configuration change, press and
release DISPLAY ADJUST NO to return to the screen with the gas
choices shown in Step 10 on page 219. You can either scroll down to
END and press POWER ENTER RESET to return to the Configure
Gases screen or continue from Step 10 on page 219 to select a new gas.

18. Use RANGE ▼ SHIFT to place the cursor next to END.

19. Press and release POWER ENTER RESET to return to the main menu.

20. Use RANGE ▼ SHIFT to place the cursor in front of NORMAL

OPERATION at the bottom of the main menu.

CHANGE TO PRO ?

PRESS YES OR NO

CONFIGURE GASES

> CAT : CH4 (CAT)

OP1 : PRO (TC)

OP2 : --- (---)

OP3 : --- (---)