94 • Calibrating Using the Single Calibration Method

EAGLE 2 Operator’s Manual

If any of the sensors cannot be adjusted to the proper value, a screen
displays that indicates a calibration failure and lists the sensor(s) that
failed to calibrate. In the example below, the oxygen and H

2

S channels

failed calibration. The other sensors calibrated normally.

The buzzer and LED arrays activate in a double pulsing pattern. Press
and release the POWER ENTER RESET button to reset the alarm and
return to the Calibration Mode Screen. Attempt to calibrate again. If the
failure continues, investigate the cause. See “Troubleshooting” on
page 70

25. Disconnect the tubing from the probe.

26. Use the RANGE ▼ SHIFT button to place the cursor next to the

NORMAL OPERATION menu option, then press and release the
POWER ENTER RESET button to return to Measuring Mode.

Calibrating Using the Single
Calibration Method

Single Calibration allows you to calibrate one channel at a time. This is
useful if you only want to calibrate one or two channels.

Setting the Fresh Air Reading

1. While in the Calibration Mode Screen, move the cursor to the

PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT
button.

FAIL SENSOR

< > <OXY>
<H2S> < >

ENTER TO CONTINUE

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION
> PERFORM AIR ADJUST

NORMAL OPERATION