EAGLE 2 Operator’s Manual

TC Calibration • 211

15. To adjust the values on the screen, press and hold the RANGE ▼ SHIFT

button, then press the DISPLAY ADJUST NO button and release both.
The following screen appears with the cursor next to CH4.

16. Place the cursor next to the channel whose gas value you want to change

using the AIR ▲ YES and RANGE ▼ SHIFT buttons.

17. Press and release the POWER ENTER RESET button to select the

channel. The calibration gas value begins to flash.

18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the

calibration gas setting to the desired value.

NOTE: The calibration gas setting value be set lower than the low alarm

setting. If the calibration gas value listed on the calibration cylinder
is lower than the current low alarm setting, enter Setup Mode and
change the low alarm setting. See “Updating the Alarm Point
Settings” on page 115 for instructions. If you need to change the
alarm point setting only to perform a calibration, make sure that you
change the alarm point setting back to its original value once the
calibration has been performed.

19. Press and release the POWER ENTER RESET button to save the change.

The calibration gas value stops flashing.

20. Repeat Step 16 through Step 19 for any other channels that need to be

changed.

21. When you are done adjusting the calibration gas values, move the cursor

down past the bottom of the screen next to END.

ADJUST AUTO

CALIBRATION VALUES

> CH4 50 %LEL

OXY 12.0 vol%
H2S 25.0 ppm

CO 50 ppm

ADJUST AUTO

CALIBRATION VALUES
> END