Bio-Rad Profinia™ Protein Purification Instrument User Manual

Page 112

Advertising
background image

9.3.4 Cleaning Sample Lines

This procedure is used to clean the two sample lines; buffer position 8 holds the 20%
ethanol solution and is used as a final system purge:

1. Touch the Data/Utilities button on the system home screen.

2. Touch the Diag/Maint Functions button.

3. On the Diag/Maint Functions screen, touch the up and down arrows to scroll through

the list of functions and highlight “Clean Sample Lines”, then touch the Select button.

4. Follow the directions displayed on the screen (Figure 9.8).

5. Touch the Start button. The system washes the sample lines with water followed by a

system purge with ethanol from buffer position 8. A progress bar displays while cleaning
is in process.

When cleaning is complete, the system returns to the Diag/Maint Functions screen.

Fig. 9.8. Prepare system to clean sample lines screen.

105