Section 13
Care and Maintenance

Proper care and maintenance of Profinia

system parts help keep the instrument operating

at optimum performance levels.

13.1 Cleaning the Instrument

The Profinia instrument automatically flushes system lines at the end of each purification
run. To keep the system clean, there are several automated cleaning and storage procedures:

Daily cleaning — buffer, sample, and fraction line cleaning

Weekly cleaning — system NaOH and ethanol washes

Storage

The procedures are selected from the Diag/Maint Functions menu, which is accessible via
the Data/Utilities menu. The procedures outlined below are designed to use water, ethanol,
or sodium hydroxide (NaOH). Other cleaning solutions, such as 0.5 M hydrochloric acid
(HCl), or 6 M guanidine HCl, may also be used. When using multiple cleaning solutions (for
example, 0.1 M NaOH followed by 0.5 M HCl), flush the system with water between steps
to prevent interactions between the cleaning chemicals.

Before running the cleaning procedures, remove the cartridges and connect the cartridge inlet
and outlet fittings by turning the plastic collars clockwise. Remove all fraction tubes and
place the cleaning tray in the fraction collection area. To run each cleaning procedure:

1. Touch the Data/Utilities button, then the Diag/Maint Functions button on the lower

toolbar to retrieve the menu for cleaning functions.

2. On the Diag/Maint Functions screen, touch the up and down arrows to scroll through

the list of functions and highlight the desired cleaning procedure.

3. Touch Select, then follow the touch screen instructions.

13.1.1 Daily Cleaning

When finished using the system for the day, perform buffer, sample, and fraction line
cleaning procedures as described in section 13.1.4.

13.1.2 Weekly Cleaning

Once a week, run the system NaOH wash, then the system ethanol wash as described in
section 13.1.4.

13.1.3 Storage

If the system will not be used for more than a week:

1. Run the system NaOH wash, then the system ethanol wash as described in section

13.1.4.

2. Empty the diluent and waste bottles.

3. Cap the ends of the buffer and sample inlet tubes and fraction outlet tubes with the

caps the instrument was shipped with (replacement Catalog #620-0444) to prevent the
entry of dust or foreign objects.