Diluent bottle — filled with deionized water

Wrenches:

Open end 7/16"

Open end 1/2"

Open end 9/16"

Replacement check valves (if required):

Inlet — catalog #620-0413

Outlet — catalog #620-0414

To access the system check valves:

1. Use the procedure in section 13.2 to remove the instrument access panel.

2. Locate the pump assembly; the inlet check valves are located at the bottom of the

pump, and the outlet check valves are located at the top (Figure 13.3).

Fig. 13.3. Location of the inlet (bottom) and outlet (top) check valves on the pump assembly.

Note: The inlet and outlet check valves look similar, but they are different parts. Assembly
and disassembly procedures for the valves are not identical; be sure to return them to their
correct locations (inlet valves on the bottom and outlet valves on the top). The instrument
will not operate properly if the check valves are replaced in the incorrect positions.

146

Outlet check
valves

Inlet check
valves