Bio-Rad Profinia™ Protein Purification Instrument User Manual

Page 79

Advertising
background image

Fig. 7.4. Enter Run & Sample Information screen.

The following method-specific information can be entered or selected:

Method name — name for the method that can be stored in the system (the system
stores up to 35 program methods) and retrieved for later use. Enter the method name
before editing and saving the method parameters

Username — optional user-designated name 3–15 characters in length to document
the operator of the run; default is “Anonymous”. The entered username remains on the
user interface until a new user edits the name or when cycling the instrument power. In
the latter case the username defaults to “Anonymous”

Note: The programmed method is saved with the date it was created and the username.

The information below the text box line on the screen relates only to the run and will not be
saved with the method. This information will be stored with the run data file.

The following run-specific information can be entered:

Run name — user-designated file name for the run, up to 30 characters in length

(External fc) — external fraction collection option; currently not available for use

Lot # tracking — optional; lot number information for the buffers, cartridges, and kits
used for the run (“BLANK” indicates no information has been entered). The word
“PRESENT” indicates that lot number information was entered. The lot number
information remains in memory as long as the instrument power remains on. When
cycling the power, the lot number information reverts to “BLANK”

The following sample-specific information can be entered or selected:

Sample name — optional; user-assigned name for the sample, up to 15 characters in
length (“UNTITLED” indicates no information has been entered). If you have chosen to
perform a run with two samples, you can assign a name for each sample