screen (Figure 8.3). See section 7.3 for details on editing the method and entering run and
sample information.

Fig. 8.3. Enter Run & Sample Information screen retrieved when saved method is selected.

The following options are available in the Enter Run & Sample Information screen:

Save — saves revised information (button displays only when edits have been made).
Touch the Save button to store the edited method to memory

Note: The Save button in this screen is programmed as a “save as” function. When the
name of a saved method is edited in the Enter Run & Sample Information screen, the
newly named method is stored in memory in addition to the original saved method. A
popup message displays, indicating method memory is full when 35 methods are stored.
You will need to delete a method prior to saving a new method. The new method can be
run without saving it to memory. When memory is full, the upper right status corner of the
screen displays “F”.

Edit — accesses the alphanumeric or numeric keypad to edit method name, username,
and run and sample information

Edit method — accesses the method steps for editing the individual step parameters

Back — returns to the Select Method Type & Options screen; allows editing of the
method options

Save — Methods — returns to the “Select saved Method System to View” screen

For details on setting up and starting the run, see section 6.

Important note for Program Methods users:

The saved methods using older version of firmware may not work properly using
newer versions of firmware. The best practice is to record the parameters saved in
the old methods and reprogram it as a new method on the new firmware. Refer to
Appendix K for Profinia planning worksheets to record parameter changes
.