Bio-Rad Profinia™ Protein Purification Instrument User Manual
Page 199
![background image](/manuals/600410/199/background.png)
Fig. 1. Typical chromatogram and SDS-PAGE for Bio-Rad 1 ml Protein A plus 10 ml desalting method.
Section 3
Frequently Asked Questions and Troubleshooting
Problem
Possible Cause
Possible Solutions
Antibody detected in 1A
Cartridge binding capacity
Load a smaller volume of starting
(or 2A) flow-through fractions
exceeded
material or increase cartridge
size so that the binding capacity
is not exceeded
Sample or equilibration buffer
Dilute sample with additional
not at neutral pH or correct ionic binding buffer or adjust pH to
strength for optimum binding
ensure neutral sample pH
Sample loading is too fast for
Choose the “Low” flow rate
`
efficient binding
option in the Protein A methods
and, if necessary, dilute sample
to reduced concentration of IgG
No antibody detected in 1D
Sample does not contain
Check starting sample on
(or 2D) elution fraction
immunoglobulin or
SDS-PAGE to ensure that
species/subclass of IgG does
antibody is present and refer to
not have affinity for Protein A
cartridge manual to confirm
specificity of species/subclasses
for Protein A resin
Antibody lacks function in
Antibody is acid labile
Increase pH of elution buffer and
downstream assay or
use Protein A Custom method to
appears degraded on
collect larger volume from elution
SDS-PAGE
fraction
Antibody is temperature
Carry out purification at 4°C and
sensitive
set Profinia instrument to
cold-room operation setting
Low purity of antibody on
Sample load and wash steps
Select “Extended” wash option
SDS-PAGE or residual
do not return to baseline with
for Protein A method during setup
contamination from
standard Protein A method
to increase wash time
starting material
wash option
192
Fraction 1A-Flow-through, 1B-Wash, 1D-Affinity-purified + Desalted IgG