Bio-Rad Profinia™ Protein Purification Instrument User Manual

Page 199

Advertising
background image

Fig. 1. Typical chromatogram and SDS-PAGE for Bio-Rad 1 ml Protein A plus 10 ml desalting method.

Section 3
Frequently Asked Questions and Troubleshooting

Problem

Possible Cause

Possible Solutions

Antibody detected in 1A

Cartridge binding capacity

Load a smaller volume of starting

(or 2A) flow-through fractions

exceeded

material or increase cartridge
size so that the binding capacity
is not exceeded

Sample or equilibration buffer

Dilute sample with additional

not at neutral pH or correct ionic binding buffer or adjust pH to
strength for optimum binding

ensure neutral sample pH

Sample loading is too fast for

Choose the “Low” flow rate

`

efficient binding

option in the Protein A methods
and, if necessary, dilute sample
to reduced concentration of IgG

No antibody detected in 1D

Sample does not contain

Check starting sample on

(or 2D) elution fraction

immunoglobulin or

SDS-PAGE to ensure that

species/subclass of IgG does

antibody is present and refer to

not have affinity for Protein A

cartridge manual to confirm
specificity of species/subclasses
for Protein A resin

Antibody lacks function in

Antibody is acid labile

Increase pH of elution buffer and

downstream assay or

use Protein A Custom method to

appears degraded on

collect larger volume from elution

SDS-PAGE

fraction

Antibody is temperature

Carry out purification at 4°C and

sensitive

set Profinia instrument to
cold-room operation setting

Low purity of antibody on

Sample load and wash steps

Select “Extended” wash option

SDS-PAGE or residual

do not return to baseline with

for Protein A method during setup

contamination from

standard Protein A method

to increase wash time

starting material

wash option

192

Fraction 1A-Flow-through, 1B-Wash, 1D-Affinity-purified + Desalted IgG

Advertising