Table 13.1. Diameters and lengths required for Profinia system tubing connections.

System Connection

System Label

Tubing OD,

Tubing

Inches

Length,

Inches

Buffer bottle to inlet select valve port

B1 to ISV 1; B2 to ISV 2;

1/8

21

B3 to ISV 3; B4 to ISV 4;
B5 to ISV 5; B6 to ISV 6;
B7 to ISV 7; B8 to ISV 8

Inlet select valve common to proportioning

ISV Com to PV (NC)

1/8

8

valve NC port

Diluent bottle inlet to proportioning valve

Water to PV (NO)

1/8

17

NO port

Proportioning valve common to inline filter

PV (COM) to PUMP IN

1/8

12

Pump inlet tee (in pump block) to inlet check

Pump in to check VLV

1/8

5

valve (below pump block), 2 each

Pump outlet tee (in pump block) to inlet check Pump out to check VLV

1/16

5

valve (above pump block), 2 each

Pump outlet to pressure sensor inlet

Pump out to pressure

1/16

8

Pump wash loop (top of pump block,

PWash loop

1/8

4

near sheet metal)

Pump wash outlet to waste tee (bottom

PWash out to waste

1/8

8

pump block)

Inline filter outlet (above buffer bottles) to

Filter out to pump in

1/8

18

pump inlet

Column valve 1 port 2 to pump wash in

CV1 P2 to PWash in

1/8

8

(bottom pump block)

Pressure sensor to column valve 1 common

Pressure to CV1 COM

1/16

8

(center) port

Column 1 tee to UV monitor 1 flow cell

Tee to UV1 IN

1/16

8

(below UV monitor)

UV monitor 1 flow cell outlet (above UV

UV1 to CV2 COM

1/16

8

monitor) to column valve 2 common

(center) port

Column valve 2 port 5 to column valve 2 tee

CV2 P5 to tee

1/16

5

UV monitor 2 flow cell outlet to column

UV2 out to tee

1/16

5

valve 2 tee

Column 2 tee to conductivity monitor inlet

Tee to COND IN

1/16

5

(or pH flow cell inlet)

(Optional) pH flow cell to outlet selector valve

pH to OSV

1/16

6

common (center) port

Sample inlet short 1/8" tube with air sensor

S1 or S2

1/8

2.5

(inside electronics cabinet)

Sample inlet 1/16" tube from union (inside

S1 to ISV P9; S2 to ISV P10

1/16

16

electronics cabinet) to inlet select valve

Column valve 1 port 5 to column 1 tee

CV1 P6 to COL1 IN

1/16

9

Column valve 1 port 6 to column 2 Inlet

CV2 P6 to COL2 IN

1/16

6

Column 1 out to column 1 tee

COL1 out to TEE

1/16

27

Column 1 tee outlet to UV monitor flow cell 1

COL1 TEE to UV1 IN

1/16

27

Column 2 outlet to UV monitor 2 inlet

COL2 out to UV2 IN

1/16

9

(under UV monitor)

Waste manifold to waste outlet

Manifold to out

1/8

12

162