Appendix d profinia, System error messages – Bio-Rad Profinia™ Protein Purification Instrument User Manual

Page 183

Advertising
background image

Appendix D
Profinia

System Error Messages

The error messages listed below are displayed on the Profinia touch screen when instrument
operation errors occur.

Message

Possible Cause(s) and Solutions

Cannot Talk to I2C device X*

Interprocessor communication problems;
possibly due to loose connections.

Check internal connections and try again.

Valve configuration error

Dirty valves, defective valves, or defective
cables.

Bad valve X*
Position error: valve X*
Can’t home: valve X*
Can’t find prop valve

Internal error: X*

Power cycle the unit; contact Bio-Rad if error
persists.

Air in line detected!

Replace or add reagent.
Check all reagents and deionized water.
Press Start to re-prime the line and continue
the method.
Press Stop to terminate run.

Pump Error

Contact Bio-Rad if error persists.

Over pressure detected @60 psi

Correct and continue.

operations suspended

Over pressure detected-paused

Correct and continue

check cartridges and method

Over pressure-did you remove

Correct and cycle power

the waste line shipping plug?

Method storage is full saving

Delete a stored method or proceed without
(35 methods can be stored; user can
continue without saving methods).

Internal hardware error, disconnected

Check connections; contact Bio-Rad if error

components, check device X*

persists.

Trouble with PV board, error X*

Check connections.

Valve move while pumping, run paused

Contact Bio-Rad if error persists.

There is a problem using your memory

If problem persists, power-cycle the unit and

stick. Please re-insert it and try again.

retry. Use a different portable memory
device.

*”X” represents an internal device address. Please contact Bio-Rad technical support (1-800-4BIORAD) for assistance.