Appendix d profinia, System error messages – Bio-Rad Profinia™ Protein Purification Instrument User Manual

Page 183

Advertising
background image

Appendix D
Profinia

™

System Error Messages

The error messages listed below are displayed on the Profinia touch screen when instrument
operation errors occur.

Message

Possible Cause(s) and Solutions

Cannot Talk to I2C device X*

Interprocessor communication problems;
possibly due to loose connections.

Check internal connections and try again.

Valve configuration error

Dirty valves, defective valves, or defective
cables.

Bad valve X*
Position error: valve X*
Can’t home: valve X*
Can’t find prop valve

Internal error: X*

Power cycle the unit; contact Bio-Rad if error
persists.

Air in line detected!

Replace or add reagent.
Check all reagents and deionized water.
Press Start to re-prime the line and continue
the method.
Press Stop to terminate run.

Pump Error

Contact Bio-Rad if error persists.

Over pressure detected @60 psi

Correct and continue.

operations suspended

Over pressure detected-paused

Correct and continue

check cartridges and method

Over pressure-did you remove

Correct and cycle power

the waste line shipping plug?

Method storage is full saving

Delete a stored method or proceed without
(35 methods can be stored; user can
continue without saving methods).

Internal hardware error, disconnected

Check connections; contact Bio-Rad if error

components, check device X*

persists.

Trouble with PV board, error X*

Check connections.

Valve move while pumping, run paused

Contact Bio-Rad if error persists.

There is a problem using your memory

If problem persists, power-cycle the unit and

stick. Please re-insert it and try again.

retry. Use a different portable memory
device.

*”X” represents an internal device address. Please contact Bio-Rad technical support (1-800-4BIORAD) for assistance.