26

5. Remove the top cap from the cartridge. Move the upper

cartridge holder down to the cartridge.

6. Twist the black collar to tighten the upper fitting onto the

cartridge until snug.

Note:
Do not overtighten fittings. These fittings are designed to seal when turned
until snug. The cartridge flange may be damaged by overtightening and may
then leak. If leaking occurs, replace the cartridge.

7. Repeat this procedure to install a cartridge in the

cartridge 2 position, if required.

8. When cartridge installation is complete, touch the Next

button on the Install Cartridges screen. You will proceed
to the Start Run screen.

Step 4: Start the Run

4A. Start Run Screen

1. To view data in real time on a computer with Profinia

software, connect the instrument to a computer, and
start Profinia software before starting the run.

2. Any externally connected reagents or samples should be

manually primed before starting the run.

3. The View Steps button allows you to view detailed steps

of the method.

4B. Run In Progress Screen

The Run In Progress screen displays progress of the run.

S1 and S2 arrow marks indicate the approximate time for
sample elution

Expected elution time is displayed; the sample can be
collected at this time

The cleaning process is shown in the progress bar as the
time portion after sample elution (S1 and S2)

Cleaning and Run Data

1. When elution is complete, proceed with instrument

cleaning as indicated on the Profinia display.

2. Run data can be downloaded to a USB portable memory

device through the USB port in the front of the instrument
after the cleaning process.