2.2 First-Time Use of the Profinia System

Before using the Profinia system, perform buffer, sample, and fraction line cleaning
procedures to flush the system and remove any air trapped in the lines.

2.2.1 Preparing the Instrument for Cleaning

Fig. 2.6.a. Profinia system ports without bottles and tubes; 2.6.b. Profinia cleaning tray placement in the
fraction collection area.

Note: For detailed instructions on installing reagent bottles, sample tubes, fraction tubes,
and using cartridge connections, see section 6.4 and 6.5.

To prepare the Profinia system for cleaning prior to its first use:

1. Remove the plastic bag covering the Profinia instrument and remove the “No Lift” stickers

from the top cover.

2. Turn the instrument around so the front faces you, and turn the system on using the

switch on the lower right in front of the diluent (water) bottle.

3. Open the instrument top cover. Remove the plastic caps from the ends of the buffer

and sample tube lines.

8

a.

b.

Buffer tray area

Cartridge
holders

Waste
bottle
holder

Sample
loading area

Diluent (water)
bottle holder

Fraction collection area