Fig. 13.13. UV monitor assembly.

3. Follow the lamp’s power cable to the metal panel, and unplug it (Figure 13.14).

Note: Depress the locking tab before pulling the connector out.

Fig. 13.14. Unplugging the UV lamp power cord.

4. Reach behind the black UV monitor assembly, and remove the two screws holding the

lamp assembly using the 9/64" Allen wrench provided with the new lamp. Pull the lamp
carrier and filter holder gently from the UV monitor body (Figure 13.15). Set the filter
holder and screws aside in a safe place.

155