6.6.3 Priming the Cartridge Lines
The Priming in Progress screen (Figure 6.21) displays a progress bar that indicates the
percentage of priming time completed. Priming is completed in 30 seconds.
Fig. 6.21. Priming in Progress screen.
If the system leaks from the cartridge positions during priming, the fittings are not
properly connected. To correct this problem:
1. Touch the Pause button in the lower right corner of the screen to stop the priming
procedure.
2. If necessary, reconnect the upper and lower cartridge fittings for each position (1 and 2).
3. Touch Start to resume the priming procedure, or touch the Back button to return to the
previous screen and restart the priming process.
When priming is finished, you will automatically proceed to the Install Cartridges screen.
6.6.4 Installing Cartridges
The Install Bio-Scale Mini Cartridges screen (Figure 6.22) shows the required Bio-Scale
™
Mini cartridges and their respective positions. It also displays instructions for installing the
cartridges into the Profinia system.