Fig. 6.19. Select cartridge status screen.

6.6.2 Setting Up Cartridge Positions for Priming

Follow the instructions on the screen to prepare both cartridge positions for priming. Both
positions need to be closed as indicated on the screen (Figure 6.20).

Note: Make sure the luer fittings are connected properly. Do not overtighten.
Overtightening can lead to stripped threads which, over time, can lead to leaking.

When both cartridge positions are closed, touch the Start button to begin the priming
process.

Fig. 6.20. Prepare for Cartridge Line Priming screen showing instructions for connecting cartridge
positions (left); upper and lower fittings connected (closed) for both cartridge positions (right).