3. On the Diag/Maint Functions screen, touch the up and down arrows to scroll through
the list of functions and highlight “Affinity Cartridge Cleaning in Place”, then touch the
Select button.
4. Select the size of the cartridge to be cleaned (1 ml or 5 ml).
5. Follow instructions displayed on the screen (Figure 9.16).
6. After the cartridge and reagents have been installed press the Start button.
7. At the end of the cleaning method the user has the option to end or repeat the method.
(Repeating the method will subject the cartridge to another cleaning cycle).
The cleaning in place method steps are listed in Table 9.1. Note that 20% aqueous ethanol
needs to be supplied at buffer position 8. This is for system cleaning only and is not part of
the cleaning in place method. The flow rate for 1 ml and 5 ml cartridges is 1 ml/min and 5
ml/min, respectively.
Table 9.1. Steps of the Affinity Cleaning in Place Method
Buffer
Buffer
Buffer
Column
Flow Rate (ml/min)
Port
Conc.
Volumes
1 ml Cartridge 5 ml Cartridge
Equilibration
B5
1x
2
1
5
Cleaning
B6
1x
10
1
5
Equilibration
B5
1x
5
1
5
Storage
B7
1x
5
1
5
20% Ethanol
B8
1x
–
1
5
Fig. 9.16. Affinity Cartridge Cleaning in Place screen.
9.3.13 Desalting Cartridge Cleaning in Place
A cleaning in place method for 10 ml and 50 ml Bio-Gel P6 desalting cartridges is available
through the Profinia Data/Utilities Menu for storage or cleaning with user defined reagents.
This method uses 1x buffer concentrates at predefined flow rates of 4 and 20 ml/min for 10 ml
and 50 ml cartridges respectively. For Profinia coldroom setting the flow rates are reduced
to 2 and 10 ml/min to prevent overpressure (Table 9.2 and Figure 9.17).
113