Running an Experiment

User Guide

| 103

Running an Experiment

Important:

When running an experiment manually, ensure that the valves are

set in the correct positions before starting the pump.

Clicking Start in the System Pump dialog box starts the run. You can change valve

positions while the pump is running and your manual run is in progress. You can

stop a run at any point and save it. You can then restart the pumps and continue the

run with a different name.

Note:

Complete the following procedures consecutively to perform a manual

run.

To select a fluidic scheme and install accessories

1.

(Optional) If your hardware configuration has changed, do one of the following:

Enter manual control mode on the touch screen.

On the computer, start ChromLab software and click Manual Run in the

Home window.

2.

Select Tools > Change Fluidic Scheme and choose the fluidic scheme that
matches the configuration of the devices connected to your NGC system.

Note:

If your NGC system includes multiple valves of the same type and

this is the first time you are using the fluidic scheme, the Fluidic Scheme

Mapping dialog box appears. Use this dialog box to map your NGC system

to the fluidic scheme.

3.

Install the accessories required to run the experiment:

a.

Connect a column.

b.

Install fraction collector racks and tubes.