Changing chromatogram view settings – Bio-Rad ChromLab™ Software User Manual

Page 40

Advertising
background image

3

| System Control

38

| NGC Chromatography Systems and ChromLab Software

Changing Chromatogram View Settings

You can change trace display attributes and other run view settings in the System

Control window. Changes you make to the following settings are saved and used in

the display of subsequent runs:

Trace Show/Hide state

Trace color

Trace y-scale range (Min/Max y values)

Active x-scale units

Show/Hide state of events, fractions, errors, and annotations

Delta Col
Pressure

Column Switching Valve

Available when at least one column switching
valve is present

Tip:

When two or more column switching

valves are present, the traces appear in
different shades of blue on the
chromatogram. The trace number
corresponds to the valve number in the fluidic
scheme. For example,

Col Pressure 1 in the

legend and the chromatogram corresponds
to C1 in the fluidic scheme.

Sample Pump
Pressure

Sample Pump

Sample pump pressure (psi or MPa)

Temperature

Single-Wavelength UV
and Multi-Wavelength
UV/Vis Detectors

The temperature of the fluid, read at the Single-
Wavelength UV or Multi-Wavelength UV/Vis
module

Flow Rate

System Pump

Current system flow rate (ml/min)

Sample Pump
Flow Rate

Sample Pump

Sample pump flow rate (ml/min)

Table 1.

Trace Definitions, continued

Trace

Module

Explanation