Bio-Rad ChromLab™ Software User Manual

Page 237

Advertising
background image

Peak Integration

User Guide

| 235

Molecular mass (kDa) — the molecular mass of the protein. By default

this field is empty. Molecular mass values that you enter in the Peaks table

are automatically populated to the Fractions table for relevant fractions.

Molecular mass values that you enter in the Fractions table are

not

automatically populated to the Peaks table.

Molarity (

M) — the calculated molar concentration of the protein for the

specified peak. This value is computed from the calculated concentration

and molecular mass.

280/260 (or 280/255) — the ratio of absorbance at 280

nm and 260 nm (or

255 nm) used to determine purity of protein for each peak in the 280 trace.

The ratio appears in the 280 nm trace section of the table. The ratio is

calculated using baseline-subtracted values of the UV trace at the

indicated retention time or volume of the 280 nm peak.

Note:

This column appears in the Peaks table after peak integration is

performed when both the 280 nm and 260 nm (or 255 nm) traces are

detected. Depending on which trace is present, the column name can be

either 280/260 or 280/255.

Table Display Order and Column Selection

You can change the order of the table columns. Column selection and order settings

are specific to the user and apply to subsequent peak integration results tables.

To change column display order

Drag columns to new locations in the Peaks table.

Showing or Hiding Columns

As in the Runs/Traces table, you can show or hide columns in the Peaks table by

choosing Show Column Chooser in the context menu.

Note:

You can show or hide columns in the table without affecting the data the

columns contain.

To hide columns in the displayed Peaks table

Click the column heading to hide and drag it out of the table.

Advertising