Bio-Rad ChromLab™ Software User Manual

Page 139

Advertising
background image

Phase Controls and Parameters

User Guide

| 137

In gradient segments, lower numbers yield increasing gradients. Higher

numbers yield negative gradients.

Load/Inject Sample — controls sample loop loading, sample injection,

and loop washing.

Fraction Collection — turns on fraction collection for the phase, sets the

fraction collection scheme, enables you to select the primary collector

(BioFrac fraction collector or outlet valve), sets the start tube or port, and

sets the fraction size.

Change Valve — changes the selected valve to the selected position.

Lamp Control — sets the wavelength and turns the lamp on/off for the

detector selected in the fluidic scheme.

Zero Baseline — sets zero as the baseline for the UV detector selected in

the fluidic scheme.

Hold Until — keeps the phase at the current composition until a button is

pressed, timeout occurs, or a threshold value is reached.

Pause — pauses the method until a button has been pressed or a timeout

time has been reached.

System Wash — sequentially switches through pumps, valves, and

sample loops and flows a defined volume of cleaning solution through

each position at a defined flow rate.

Note:

This step is not supported when either Gradient Segments or Load

Inject Sample steps are included in the method.

Advertising