System settings, Delay volume tab, Customize system settings. see – Bio-Rad ChromLab™ Software User Manual

Page 71

Advertising
background image

System Settings

User Guide

| 69

System Settings

The System Settings dialog box enables you to customize your system. From this

dialog box you can

Set the delay volume

Control the flow rate to prevent overpressure

Enable remote access to ChromLab from an iPad, Android device, or

another PC

Reset the trace settings, the minimum and maximum y-scale values, and

the trace colors in the chromatogram to their factory default settings

Enable or disable external detectors that are connected to the NGC

instrument via a SIM device

Enable or disable the ability to send UV signals to external devices that are

connected to the NGC instrument via a SIM device

Set air sensors to detect either end of sample or end of buffer

Enable the system to send email messages about system events to a list of

users

Set a unique name for the NGC system

Note:

This dialog box is accessible in manual mode. The settings are saved

and the fraction collector is also synchronized in method runs.

Delay Volume Tab

Delay volume is the volume of plumbing between the UV detector and the fraction

collector drophead and/or the outlet valve. It indicates the delay between the

detection of the fraction’s UV peak and the time it takes for this peak to traverse the

volume of tubing and flow cells to the fraction tube into which it is collected.