Bio-Rad ChromLab™ Software User Manual

Page 295

Advertising
background image

Performing Multicolumn Purification Chromatography

User Guide

| 293

Multicolumn Tandem 2-D Affinity Templates

Tasks

1.

Insert a second column switching valve (referred to as the loop
valve) near the sample inject valve on the NGC instrument and map
as C2

2.

Short length of tubing from port F on the sample inject valve to the
left inlet port on the loop valve (C2)

3.

Short length of tubing from port E on the sample inject valve to the
right inlet port on the loop valve (C2)

4.

One 5 ml loop tubing for each affinity column to the top and bottom
ports on the loop valve (C2)

For example, if you have two affinity columns

Attach loop tubing to Column Top port 1 and Column Bottom
port 1

Attach loop tubing to Column Top port 2 and Column Bottom
port 2

5.

Tubing from port 2 on the outlet valve to the syringe port on the
sample inject valve

6.

Affinity columns at ports 1–4 on C1

7.

Desalting or SEC column at port 5 on C1

Multicolumn Tandem Affinity Templates

Tasks

1.

Insert a second column switching valve (C2) into the NGC
instrument

2.

Short length of tubing from outlet port on C1 to the inlet port on C2

3.

Short length of tubing from outlet port on C2 to the UV detector

4.

Affinity columns at ports 1–5 on C1

5.

Desalting column at port 5 on C2