Flow buffer selection – Bio-Rad ChromLab™ Software User Manual

Page 135

Advertising
background image

Method Settings Parameters

User Guide

| 133

BioFrac Settings — by default, fraction collection begins in the first tube of the

first rack. The fraction size is the default for the rack type.

BioFrac Collection Pattern — by default, Serpentine is selected. For

microplates, fractions can also be collected by rows or columns.

Outlet Valve Settings — by default, fraction collection begins from

Outlet valve 1 (O1) Port 2. O1 Port 1 connects to the BioFrac™ fraction

collector if it is present or directs flow to waste if the fraction collector is absent.

The default fraction size is 50 ml.

Flow

Flow Rate — shows the default rate of flow for the method.

Control the flow to avoid overpressure — monitors the pressure and reduces

the flow rate by 50% when the pressure gets within a certain percentage of the

maximum (default is 80%, see

Control Flow Tab

in Chapter 3 for more

information).

Detect end of buffer with air sensor — stops the system pumps if air is

detected in any of the lines connected to air sensors set to detect end of buffer.

Air sensor settings are determined on the Air Sensors tab in the System

Settings dialog box. (See

Air Sensors Tab

in Chapter 3 for more information).

Buffer Selection

Note:

When the configuration does not include a buffer blending valve, buffer

settings are inactive.

You can choose one of three Buffer Selection modes:

Manually Prepared — buffers are user prepared and fed directly into the

pumps or selected by means of buffer inlet valves.

Manually Prepared via Blending Valve — buffers are user prepared and

fed to the pump through buffer blending valve ports Q1 and Q4. Gradients

are formed by the blending valve rather than the pumps. Because of this,

both pumps can run at their full capacity simultaneously, which doubles

the available flow rate range.

Advertising