Bio-Rad ChromLab™ Software User Manual

Page 72

Advertising
background image

3

| System Control

70

| NGC Chromatography Systems and ChromLab Software

When the delay volume is set, the fraction collector or outlet valves wait for the

fraction peak to travel from the detector to the drophead (or port) before collection

begins.

Fraction collection on the instrument can be delayed relative to the

collection information shown in the chromatogram and status panels. When the run

reaches the end of rack or the last port on the outlet valve, the chromatogram and

status panel might display the start tube and rack or new outlet port before

collection actually starts.

Note:

If an outlet valve is included in the fluidic scheme and you change the

collection port or fraction size during the run, the first container in the new

scheme collects the designated fraction size as well as the indicated delay

volume. All succeeding fraction containers collect only the designated fraction

size.

To set the delay volume

1.

Select File > System Settings to open the System Settings dialog box.

2.

Choose the Delay Volume tab.