Column selection – Bio-Rad ChromLab™ Software User Manual

Page 128

Advertising
background image

5

| Method Editor

126

| NGC Chromatography Systems and ChromLab Software

Column Selection

Under Column Selection you select the column and the column parameters for the

method. The column’s volume and pressure parameters are set automatically when

you select a column type from the list of predefined columns.

If you have one or more column switching valves in the fluidic scheme, you can

choose whether to set the parameters for a single column or for multiple columns.

When multiple columns are used in a method, the method’s phases use the

pressure parameters for any columns that are in line for that phase. You can use

method steps to place two or three columns within the same method. In these

cases, a high-pressure event is triggered when the delta pressure for one of the

column switching valves reaches the high-pressure limit as defined in Method

Settings for the columns that are in line.

You can configure the properties of multiple columns, one for each port on each

column switching valve, and add them to the method using the Configure Ports

dialog box. One column is set as default when you select its checkbox. The method

uses the default's value for column volume throughout the method regardless of any

other columns the method uses. In addition, the method uses the default's

maximum pre- and delta-column pressure until other columns are switched in line. If

two or more columns are switched in line, the method uses the maximum

pre-column pressure of the first column.

Advertising