Air Sensors Tab

The system supports up to two air sensor modules and up to eight air sensors. Each

air sensor module can contain up to four air sensors. The system detects the

number of air sensors that are attached to the instrument.

Air sensors can be set to detect air, indicating end of buffer, in lines connected to

the system pumps. When the air sensor detects end of buffer, the system stops the

system pumps and the run. You can continue the run after replenishing the buffer

and purging the lines.

Air sensors can also be set to detect air, indicating end of sample, in lines

connected to the sample pump. When the sample pump is used to load sample in

manual mode and the air sensor detects end of sample, the system stops the

sample pump to keep it from pumping air onto the column. When the air sensor

detects end of sample during the sample application phase, the sample pump

stops. In this case the run proceeds to the next phase.

To activate air sensors

1.

Select File > System Settings to open the System Settings dialog box.

2.

Choose the Air Sensors tab.

By default, only the air sensors that the system detects are enabled. The other

sensors are disabled.