Elution parameters – Bio-Rad ChromLab™ Software User Manual

Page 146

Advertising
background image

5

| Method Editor

144

| NGC Chromatography Systems and ChromLab Software

Although it is typically used after the sample application or elution phase, this phase

can be used on its own or at any point that column washing is wanted.

Elution Parameters

In this phase, the sample is eluted from the column using a gradient or isocratic flow

at defined salt concentrations (%B) over a defined volume, column volume, or time.

You can edit the parameters in the gradient segment or add more isocratic or

gradient steps to this segment. This phase can use the flow rate defined globally in

the Method Settings phase or a flow rate different from that of the other phases.

Fraction collection is enabled by default during this phase and offers options to

collect all, to collect when a certain threshold of UV or conductivity has been

reached, or within defined collection windows. When collecting fractions by

threshold, you can opt to collect the fractions that fall outside the defined threshold

range using fraction volumes different from the volumes set for the threshold

fractions.

You can choose an elution technique from the following options:

Isocratic – define the length of elution (CV, cm/hr, time, volume) and buffer

composition (%B).

Gradient – define a linear gradient (%B; default = 0–100%) and gradient

slope/duration (CV, time, volume). In this case, the pumps stop at the end

of the set duration.

Advertising