Bio-Rad ChromLab™ Software User Manual

Page 186

Advertising
background image

6

| Creating a Method

184

| NGC Chromatography Systems and ChromLab Software

checkbox is also cleared. The other steps remain selected and will be

included in the scout.

Page 3: Generate Scout Sequence

On this page you set the number of runs to be performed, name each run, and set

the parameters for the scouted steps.

Number of runs — sets the number of runs to be performed as part of the

scout experiment.

Starting Value — sets the starting value for the parameter chosen to be

scouted. The default is the starting value in the base method.

Increment Value — sets the scout run increment values. The scout

increment value can be positive or negative.

The scout sequence table comprises the following columns:

Run # — the scout run number. This field is not editable.