Calibrating pressure settings – Bio-Rad ChromLab™ Software User Manual

Page 68

Advertising
background image

3

| System Control

66

| NGC Chromatography Systems and ChromLab Software

Calibrating Pressure Settings

WARNING!

You must remove the tubing from the pressure sensor before you

calibrate it. The sensor must be at atmospheric pressure when you start.

You can calibrate sample pump pressure, system pressure, and column switching

valve pressure to zero. If pressure is not reading zero or close to it, you can reset the

pressure values to zero.

To calibrate a pressure setting

1.

Select a pressure module in the Calibrate dropdown list at the top of the
Calibration dialog box.

2.

Follow the dialog box instructions to reset the pressure setting to zero.

Note:

To calibrate pressures at the upper limit, call Bio-Rad Support.