Working with fluidic schemes – Bio-Rad ChromLab™ Software User Manual

Page 59

Advertising
background image

Fluidic Scheme Pane

User Guide

| 57

Working with Fluidic Schemes

Five sections of the fluidic scheme can be configured based on the available

modules and experiment requirements.

Pumps

Sample inject

Columns

Detectors

Sample collection

To change a fluidic scheme

1.

Do one of the following:

In the System Control window, select Tools > Change Fluidic Scheme.

On the touch screen, choose Change Fluidic Scheme on the dropdown

menu.

The Fluidic Scheme Selector dialog box appears. A descriptive tooltip appears

when you pause the pointer on a module in the fluidic scheme.

Two outlet valves

For use as a fraction collector, can
collect up to 22 large-volume fractions
when two outlet valves are daisy-
chained together.

Select outlet ports can be plumbed to
divert a fraction eluted from a column
onto a sample/storage loop to be
subsequently reinjected onto a second
column during tandem applications.

Table 3.

Fluidic Scheme Configuration Options, continued