Bio-Rad ChromLab™ Software User Manual

Page 193

Advertising
background image

Scouting on Parameters within a Method

User Guide

| 191

Generating Column Scout Sequences

This option is available only when the fluidic scheme includes the column

switching valve.

If you chose Custom as the column type and changed the default values

for column volume, maximum precolumn pressure, and/or maximum delta-

column pressure, your custom settings must be within the system limits

defined by the method settings. If your custom values are outside of this

range, a red box appears around the number. The wizard prevents you

from continuing until you enter a valid number.

Generating Sample Scout Sequences

This option is available only when the fluidic scheme includes at least one

sample inlet valve and you chose to load the loop through the sample

pump or inject sample directly onto the column in the Sample Application

phase. It is not available if you chose to load the loop manually.

If your method contains multiple Sample Application phases, scouting is

performed only on the first sample loading step in the method.

If you select Load Loop with Sample Pump in the method, the injection

volume in the scout refers to the system pump injection volume, the initial

injection volume is the same as the system pump injection volume set in

the method, and the minimum injection volume is 0.01 ml.

If you select Inject Sample on Column with Sample Pump in the method,

the injection volume in the scout refers to the sample pump injection

volume, the initial injection volume is the same as the sample pump

injection volume set in the method, and the minimum injection volume is

0.01 ml.

The maximum injection volume is 1,000 L.