Calibrations

The NGC instrument arrives factory calibrated. For the most part, the instrument will

not require further calibration. The pH probe, however, should be recalibrated each

day the instrument is used and when there are large fluctuations in temperature

because pH probe settings drift with time.

In addition to the pH probe, ChromLab includes settings for calibrating the

following:

Pump flow rate

Sample pump pressure

System pressure

Column switching valve pressure

Conductivity monitor

Calibrating these modules is optional.

Calibrating a Module

Note:

You cannot run a method during calibration.

On the Tools menu, clicking Calibrate displays a Calibration dialog box in which you

can select a module to calibrate. The System Control window also appears with the

title Calibration in Progress in the title bar.

To select a module to calibrate

Select a module in the Calibrate dropdown list at the top of the Calibration
dialog box.

Instructions for calibrating the module you selected appear in the dialog box.