Special considerations – Bio-Rad ChromLab™ Software User Manual

Page 192

Advertising
background image

6

| Creating a Method

190

| NGC Chromatography Systems and ChromLab Software

Special Considerations

Rules for generating scout sequences can vary depending on the parameter chosen

to scout. This section provides a list of considerations to take into account when

generating a scout sequence.

Generating %B Scout Sequences

You can scout on two or more steps with the same starting %B values. If

you select two or more steps with different starting values, you must clear

the steps with the different values.

When a linear gradient is selected for scouting and the value for the next or

previous %B step is the same as the selected gradient, that step is

automatically selected for scouting. Clear that step if it is not required.

Note:

If an isocratic step is selected for scouting, no other steps are

automatically selected. You can choose others manually.

When the starting value for %B is 100, the increments (on Page 3) default

to a negative value.

Generating pH Scout Sequences

This option is available only when the fluidic scheme includes the buffer

blending valve and buffer blending is selected in Method settings.

The minimum and maximum pH scouting range is defined by the buffer

selection made in Method settings.

Generating Duration Scout Sequences

Choosing two or more steps that have different starting values or that have

different units is not supported. Clear the steps with different starting

values or units to proceed.

Advertising