Annotating the chromatogram – Bio-Rad ChromLab™ Software User Manual

Page 44

Advertising
background image

3

| System Control

42

| NGC Chromatography Systems and ChromLab Software

When a single UV absorbance trace is monitored with a single-wave UV you can

Autoscale all UV traces to the same scale

.

The y-axis scale is based on the signal intensity of the UV trace with the highest

intensity and all other traces are normalized to the intensity of that trace.

Disable autoscaling

.

To autoscale the UV trace

Do one of the following:

Click Autoscale on the toolbar to choose a status.

Right-click the chromatogram, choose Autoscale UV trace, and select

a status.

Annotating the Chromatogram

You can add notes to the chromatogram in manual mode at any time during a run or

after a run completes to associate observations with data points on the

chromatogram. The annotation dialog box contains three fields: Location (in time,

volume, or CV), Title, and Description. When the annotation is saved, its title

appears at the specified location on the x-axis. The description appears in the

run

report.

You can add annotations when a method run is in progress. However, if the run ends

while the annotation dialog box is open, the annotation is saved and the dialog box

closes automatically. When a method run is complete, annotations cannot be added

to the chromatogram in System Control. You can annotate completed runs by

opening them in Evaluation mode.