System cip (clean in place) — storage parameters – Bio-Rad ChromLab™ Software User Manual

Page 147

Advertising
background image

Phase Controls and Parameters

User Guide

| 145

System CIP (Clean in Place) — Storage Parameters

Note:

System CIP must be run as the only type of phase in the method. You

can add several System CIP phases to the method if multiple cleaning solutions

are required. Bio-Rad recommends that you save System CIP as a separate

method.

If the system has been exposed to hazardous biological material, run System CIP

and Column CIP to flush the entire system tubing with sanitizing solution (for

example, NaOH) followed by neutral buffer and finally distilled water before service

or maintenance.

This phase cleans the system after purification runs by rinsing the system with

cleaning solution. Perform System CIP when required; for example, between runs

where different samples and buffers are used or before storing the system. This

helps prevent cross-contamination between runs and prevents bacterial growth in

the instrument during storage.

Depending on the modules in the fluidic scheme, choose the scope of cleaning by

selecting checkboxes for the different valves, ports, pumps, column positions,

modules, and loops to clean and fill with cleaning solution. The System CIP method

includes two System CIP phases to facilitate the use of two different solutions; for

example, NaOH, buffer solution, or distilled water. Select values for flow rate,

volume per position, and incubation time.

Note:

Each phase uses one cleaning solution. All inlet lines selected in one

phase should be immersed in the same cleaning solution.

Special Considerations

When the fluidic scheme includes a buffer blending valve, NGC systems perform

System CIP differently depending on whether the buffer selection is manually

prepared via the buffer blending valve, blended through the buffer blending valve, or

manually prepared through buffer inlet valves. Note the following special

considerations regarding System CIP with a buffer blending valve in the fluidic

scheme.

Advertising