Phase controls and parameters – Bio-Rad ChromLab™ Software User Manual

Page 138

Advertising
background image

5

| Method Editor

136

| NGC Chromatography Systems and ChromLab Software

Available Range — available ranges for the recipe concentration and pH values

vary depending on the recipe you choose. The ranges provided are

recommendations. You can choose values above the upper end of the range, but

doing so might require other adjustments to ensure accuracy. For most recipes you

can change the buffer concentration by reducing the stock concentration by 1/2 to

1/8 on Conjugate Acid/Base Pair titration and up to 1/4 on Acid or Base titration.

Stock concentration for most recipes is 0.2 M.

Tris and Phosphate are the most commonly used recipes, depending on the type of

chromatography being performed.

pH Correction — used to correct the observed pH so it matches the pH set for the

run (Desired pH) in the System Pump dialog box or Method Settings phase. pH

correction uses the difference between the desired pH and observed pH to make

the correction.

Tip:

Calibrate the pH probe regularly for accurate pH reading.

Phase Controls and Parameters

The Method Editor includes several standard phases. See

Standard Phases on

page

138

for details.

A phase is a method segment composed of a discrete block of steps created to

accomplish a specific task. Each phase comprises controls and parameters specific

to the task it is designed to accomplish.

The controls and parameters used in individual phases are as follows:

Gradient Segments — determine the length and buffer composition (%B)

for isocratic or gradient steps.

Tip:

Each phase runs at a single flow rate. To change the flow rate, select

a different phase.

In isocratic steps, initial %B and final %B must be equal. In linear segments,

initial %B and final %B are independent of each other.