Bio-Rad ChromLab™ Software User Manual

Page 236

Advertising
background image

7

| Evaluating Results

234

| NGC Chromatography Systems and ChromLab Software

Peak asymmetry — the asymmetry factor, defined as the distance from

the center line of the peak to the back slope divided by the distance from

the center line of the peak to the front slope. All measurements are made at

10% of the maximum peak height.

Note:

The asymmetry factor of a peak is usually similar to the tailing factor

of the same peak, but the two values cannot be directly converted.

Fractions — all fractions that reside between start and end. Multiple

fractions are displayed first to last.

Peak type — the integration type for a peak.

Automatic — the peak was determined automatically and has not been

manually adjusted.

Manual — the peak has been manually adjusted or added.

Path length (cm) — the path length of the UV detector’s flow cell. The

default length is 0.5 cm. The NGC system has three optional flow cells,

which are used to calculate the protein concentration: 0.2, 0.5, and 1.0 cm.

This field can be modified to display the value of the flow cell in use. When

the value in one row is changed the cell flow path length for all peaks in the

same run is updated accordingly.

Extinction coefficient ((mg/ml)

-1

cm

-1

) — the extinction coefficient of the

protein, used to calculate the protein concentration. By default this field is

empty. You can enter a value up to three significant digits. Coefficient

values entered in the Peaks table automatically populate the extinction

coefficient values for relevant fractions in the Fractions table. Coefficient

values entered in the Fractions table are

not automatically populated to the

Peaks table.

Concentration (mg/ml) — the calculated concentration of the protein for

the specified peak. This value is based on the following calculation:

(peak area/peak volume) / (extinction coefficient x path length x 1,000)

Advertising