Preparing the system, Priming and purging the system – Bio-Rad ChromLab™ Software User Manual

Page 96

Advertising
background image

4

| Performing a Manual Run

94

| NGC Chromatography Systems and ChromLab Software

Preparing the System

System preparation consists of priming the system to remove air from the pumps,

tubing, and valves followed by flushing the fluidics system to remove storage buffer,

cleaning solutions, or old buffer. Priming is performed in manual mode, while both

purging and cleaning tasks can be performed either manually or through

preprogrammed method phases. See

Chapter 5, Method Editor

for more

information about working with methods.

Priming and Purging the System

Perform the following procedures consecutively to prime and purge your NGC

systems.

Important:

Whenever you add or remove a module or change or upgrade your

NGC system hardware configuration, you must replumb and reprime the

system.

To prime the system

1.

Power on the NGC system and its connected computer.

2.

To enter manual mode, do one of the following:

On the computer, click Manual Run in the ChromLab Home window or

click the System Control tab and then click in the fluidic scheme.

On the touch screen, click Enter Manual Mode if the application is not

already in manual mode.

3.

(Optional) If your hardware configuration has changed, select a fluidic scheme
that matches your current system configuration.

a.

Select Tools > Change Fluidic Scheme.

b.

Select the appropriate fluidic scheme in the Fluidic Scheme Selector pane.

c.

Click Select to choose the manual mode fluidic scheme.

4.

Insert appropriate inlet lines into the buffer, storage solution, and/or cleaning
solutions to be primed and flushed through the system.